主要技术指标 |
质量轴稳定性Mass stability < 0.1 Da in 24 h;0.02% from 2,000 - 2,250 m/z;0.01% from 1,000 - 2,000 m/z
质量准确度Mass accuracy 0.1 Da from 5 - 1,000 m/z
分辨率 Mass resolution (manual tune) Full width at half maximum 0.5 Da;Mass resolution (autotune) Full width at half maximum 0.7 Da
正负极性极性转换时间:30毫秒
m/z范围:5 - 2,250
灵敏度:1pg利血平 S/N >500:1 MS/MS sensitivity 1 pg of reserpine injected on column Signal-to-noise ratio > 500:1
最小驻留时间:1毫秒 Minimum MRM dwell time 1 ms
MRM转换:每个时间段可做500个MRM通道;每个方法可包括>40,000个MRM通道 ?MRM transitions 500 per time segment,> 40,000 ion transitions per method
动态MRM通道:每个方法 4,000个离子转换 Dynamic MRM transitions 4,000 ion transitions per method
碰撞池离子容忍度<1毫秒 ?Collision cell ion clearance < 1 ms
?MRM灵敏度信噪比,ESI+,1pg利血平on column进样,MRM通道m/z 609-195,S/N>30,000:1
MRM灵敏度信噪比,ESI-,1pg氯霉素on column进样,MRM通道m/z 321-152,S/N>10,000:1
MRM灵敏度仪器检出限IDL,ESI+,20fg利血平on column进样,MRM通道m/z 609-195,IDL<12.5fg
MRM灵敏度仪器检出限IDL,ESI-,20fg氯霉素on column进样,MRM通道m/z 321-152,IDL<12.5fg
m/z范围:5-3000
极性转换时间:30毫秒
质量分辨率(自动调谐):0.7 Da
质量分辨率(手动调谐):0.5 Da
|
功能/应用范围 |
检测土壤污染,特别是评估人、动物和植物暴露于的土壤环境,并且尝试降低这种长期暴露,是必须进行的[3]??。
气相色谱?(GC)和液相色谱?(LC)配备质谱(MS)被广泛应用于土壤检测和分析。特别是液相色谱配备三重四级杆质谱仪(LC/MS/MS),为土壤样品中的中等极性、极性和离子型化合物的痕量分析提供了很多优势。一些LC/MS/MS土壤分析具体应用包括:
?鉴定和定量有目标多残留物质和一般未知物质,低至ng/L的水平
?高选择性,保证的污染物识别的可靠性
?最少的样品制备,保证快速的分析方法和结果的转换
?为污染物检测提供高的灵敏度,减少化合物衍生化的需求
?土壤有机物检测,提供最多的相关数据结果[3]?
|
服务领域 |
,农/林/牧/渔,轻工/纺织,石油/石化,食品/烟草,地质/矿产,矿业/冶金,钢铁/有色金属,化工/化纤,水文气象,橡胶/塑料(材料),机械制造,医疗/卫生,地质勘探,电气工程,仪器/仪表,航空/航天,环保/水利/气象/天文,生物/医药 |
技术特色 |
|